PhysioTel Digital Implant Specifications



 Feature  L series  M series
Implant Capacity1
per System
24 North America
18 Europe, China
12 Japan
 Implant Capacity
per Study Area1
48 North America
30 Europe, China
18 Japan
 Range  3 – 5 meters
 Battery On Time  L11: 105 days
L21:   84 days
L04: 95 days2
L03: 90 days2
L11R:  120 days3
 M11:  35 days
M01: 40 days
M1G: 48 days
M10: 55 days
    M00: 100 days
 Implant Weight  56 grams  13.7 grams
 Implant Dimensions  59 x 38 x 15 mm  40 x 26 x 14 mm
Activation Magnetic or via software
 Warranted Implant Duration  18 months  6 months
 Telemetry Pressure Range  -20 - 300 mmHg
 Ambient Pressure Range  670 - 800 mmHg
 Pressure Accuracy (initial) +2 mmHg/-4 mmHg 
Pressure Drift <2 mmHg/mo. (max)
0.25 mmHg/mo. (average) 
Biopotential Input Range Programmable 
(2.5, 5.0, 10.0, 20.0 mV)
Fixed 
(10.0 mV)
Biopotential Bandwidth L21, L11, and M series
ECG & EMG: 0.1-100 Hz
EEG: 3-100 Hz
L03
All Signal Types
 Ch. 1-3: 0.5-100 Hz
 L04
All Signal Types
Ch.1-3: 0.5-100 Hz; Ch. 4: 0.5-50 Hz
 Temperature Range 32 – 41 °C
 Temperature Accuracy ± 0.1 °C 
 Temperature Drift  <0.5°C
1 Note that implant model may affect maximum system and study area capacities
Warranted battery life for China L03 is 85 days and L04 is 90 days
Warranted battery life for LV capable L11R is 105 days

PhysioTel Digital case studies: