Pulmonary Compliance and Resistance Analysis Module
The Pulmonary Compliance and Resistance (PCR/PCRP) Analysis Module analyzes pulmonary flow and pulmonary pressure signals derived from either a plethysmograph chamber or a pneumotach, along with pulmonary pressure. Values for these derived signals are calculated on a breath-to-breath basis for each respiratory cycle. The graph above displays a typical pulmonary flow signal and pulmonary pressure signal with the digitally integrated volume signal. Validation marks and their explanations are also displayed. The validation marks provide visual, real-time verification of the accuracy of the system.
The list below describes the parameters calculated by the analysis module either in real-time or during subsequent analysis.
Model PNM-PCR100W
Name | | Definition |
Num |
| The number of the respiratory cycle. |
PIF, PEF |
| Peak Inspiratory Flow, Peak Expiratory Flow. |
TV, MV |
| Tidal Volume, Minute Volume |
BPM |
| The number of breaths-per-minute is calculated on a breath-to-breath basis. |
IT, ET, TT, AT | | Inspiratory Time, Expiratory Time, Total Time, Apnea Time. |
Cdyn, Res, Cond |
| Compliance, Resistance, Conductance. |
ResI1, ResI2, ResE1, ResE2 |
| Resistance calculated at various points of inspiration and expiration. |
CAvg, RAvg |
| Compliance and Resistance value averaged. |
PInspS(P1), PExpS(P2), Piso1(P3), Piso2(P4) |
| Pressure value at different time points of flow signal. |
Fiso1(F1), Fiso2(F2) |
| Flow on clearing isovolumic levels during inspiration and expiration. |
PEnh |
| Enhanced Pause. Calculated as: ((ET+AT)/RT-1)*(PEF/PIF). Formula from Noninvasive Measurement of Airway Responsiveness in Allergic Mice Using Barometric Plethysomgraphy Hamelmann et al. |
RT | | Relaxation Time. |
WOBi, WOBe, WOB |
| Work of Breathing Inspiration, Work of Breathing Expiration and Work of Breathing (sum) |
POBi, POBe, POB |
| Power of Breathing Inspiration, Power of Breathing Expiration and Power of Breathing (rate at which work was performed over inspiration and expiration). |
TVe |
| The difference between the volume at the start of expiration mark and the volume at the point prior to the next cycles start inspiration mark. |
AVol |
| Accumulated volume. |
IF50 |
| Inspiratory flow value at the point where the volume signal rises to 50% of the tidal volume. |
EF50 |
| Expiratory flow value at the point where the volume signal drops to 50% of the tidal volume. |