PhysioTel® HD Implant Specifications |
Model | Weight
(gm) | Volume
(cc) | Minimum Animal Weight | Maximum Cage Size* | Battery Life Warranted (Months) |
HD-X10 | 2.2 | 1.4 | 19 grams (sub-cu)
23 grams (IP carotid cannulation)
27 grams (IP aorta cannulation) | 33 x 33 x 14 cm | 1.5 |
HD-X11 | 2.2 | 1.4 | 19 grams (sub-cu) 23 grams (IP carotid cannulation) 27 grams (IP aorta cannulation) | 33 x 33 x 14 cm | 1 |
HD-S11 | 8 | 5.9 | 175 grams | 42 x 42 x 18 cm | 2 |
HD-S21 | 8 | 5.9 | 175
grams | 42 x 42 x 18
cm | 2 |
HD-S10 | 4.4 | 3.1 | 175
grams | 42 x 42 x 18
cm | 5 |
Parameters Measured | Pressure, Biopotential, Temperature, Activity |
Outer Material | Silicone Elastomer |
Temperature Operating Range | 34 to 41 degrees Celsius |
Shelf Life | Battery - 1 year
Sterility - 1 year |
ON-Off Mechanism | Magnetically Actuated |
Receiver Compatibility | Any PhysioTel® RLA, RPC-1 or RMC-1 |
Pressure Measurement Specifications |
Pressure Range | -20 to 300 mmHg |
Catheter Length Options | HD-S models: 8, 10 or 15 cm
HD-X11: 5 or 7 cm |
Ambient Pressure Range | 670 to 800 mmHg |
| |
Pressure Accuracy | |
| |
| |
Initial Accuracy | ± 3mmHg |
Pressure Drift | <2 mmHg/month (maximum); -0.25mmHg/month (Average) |
Frequency Response | HD-S21: 8cm catheter: >100Hz
HD-X11: 5cm catheter: >60Hz |
Biopotential Channel Specifications | HD-S21 | HD-X11 |
Input Voltage Range | ±5mV | ± 2.5 mV |
Channel Bandwidth | 0.1-145 Hz | 0.1 - 200 Hz |
Temperature Accuracy | |
Initial Accuracy | 0.1 degree Celsius |
| |
Resolution | 0.05 degree Celsius |
Temperature Drift | First two weeks = .1 degree Celsius
Per 6 months after = .1 degree Celsius |
†Recommended catheter length for small animal LVP applications.
PhysioTel® TA Implant Specifications |
Model | Weight
(gm) | Volume
(cc) | Minimum Animal Weight | Maximum Cage Size* | Battery Life Warranted (Months) |
TA-F10 | 1.6 | 1.1 | 17 g (sub-cu)
20g (IP) | 33 x 33 x 14
cm | 6 |
TA-F40 | 7.25 | 3.5 | 175
grams | 42 x 42 x 18
cm | 12 |
TA-D70 | 40 | 25 | 2.5
kilograms | 1 x 1 x 1
meter | 11 |
Parameters Measured | Temperature and Activity |
Outer Material | Silicone Elastomer |
Temperature Operating Range | 34 to 41 degrees Celsius |
Shelf Life | Battery - 1 year
Sterility - 1 year |
ON-Off Mechanism | Magnetically Actuated |
Receiver Compatibility | Any PhysioTel® RLA, RPC-1 or RMC-1 |
| |
Temperature Measurement Specifications |
Initial Accuracy | 0.15 degree Celsius (TA-F10), 0.1 degree Celsius (all other devices) |
Resolution | 0.05 degree Celsius (TA-F10), 0.01 degree Celsius (all other devices) |
Max Drift Over Warranted Battery Life | 0.05 degree Celsius first month after turn on (TA-F10)
0.05 degree Celsius first two weeks after turn on
(all other devices)
0.05 degree Celsius per six months thereafter
(all devices) |
PhysioTel® PA Implant Specifications |
Model | Weight (gm) | Volume (cc) | Minimum Animal Weight | Maximum Cage Size* | Battery Life Warranted (Months) |
PA-C10 | 1.4 | 1.1 | 17 grams
(sub-cu)
27 grams (IP) | 33 x 33 x 14
cm | 1.5 |
PA-C40 | 7.8 | 4.5 | 175
grams | 42 x 42 x 18
cm | 4 |
PA-D70 | 37 | 25 | 2.5
kilograms | 1 x 1 x 1
meter | 4 |
Parameters Measured | Pressure and Activity |
Outer Material | Silicone Elastomer
(Excluding the PA-C10) |
Temperature Operating Range | 34 to 41 degrees Celsius |
Shelf Life | Battery - 1 year
Sterility - 1 year |
ON-Off Mechanism | Magnetically Actuated |
Receiver Compatibility | Any PhysioTel® RLA, RPC-1 or RMC-1 |
| |
Measurement Specifications |
Pressure Range | -20 to 300 mmHg |
Ambient Pressure Range | 670 to 800 mmHg |
Maximum Altitude | 625 meters |
| |
Pressure Accuracy |
Initial Accuracy | ±3 mmHg |
| |
Pressure Drift |
PA-C10 | < 2 mmHg/month (Max); -0.25mmHg/month (Average) |
PA-C40 | First month < 5 mmHg
Second month < 4 mmHg
Thereafter < 3 mmHg per month |
PA-D70 | < 2 mmHg per month nominal |
PhysioTel® EA, CA, ETA and CTA Implant Specifications |
Model | Weight
(gm) | Volume
(cc) | Minimum Animal Weight | Maximum Cage Size* | Battery Life Warranted (Months) |
ETA-F10 | 1.6 | 1.1 | 17 grams (SQ)
20 grams (IP) | 33 x 33 x 14
cm | 2 |
EA-F20 | 3.9 | 1.9 | 20
grams (SQ)
25 grams (IP) | 33 x 33 x 14
cm | 4 |
ETA-F20 | 3.9 | 1.9 | 20
grams (SQ)
25 grams (IP) | 33 x 33 x 14
cm | 4 |
CA-F40 | 8.0 | 4.8 | 175
grams | 42 x 42 x 18
cm | 6 |
CTA-F40 | 8.0 | 4.8 | 175
grams | 42 x 42 x 18
cm | 6 |
CTA-D70 | 30.0 | 20.0 | 2.5
kilograms | 1 x 1 x 1
meter | 4 |
Parameters Measured | Biopotential, Temperature, Activity |
Outer Material | Silicone Elastomer |
Temperature Operating Range | 34 to 41 degrees Celsius |
Shelf Life | Battery - 1 year
Sterility - 1 year |
ON-Off Mechanism | Magnetically Actuated |
Receiver Compatibility | Any PhysioTel® RLA, RPC-1 or RMC-1 |
Biopotential Channel Specifications | ETA, EA Devices | CTA, CA Devices | Input Voltage Range | ±2.5mV | ±10mV | Channel Bandwidth | 1-200 Hz | 1 - 200 Hz | |
Temperature Measurement Specifications |
Initial Accuracy | 0.1 degree Celsius |
Resolution | 0.01 degree Celsius |
ETA-F10 Initial Accuracy | 0.25 degrees Celsius |
ETA-F10 Resolution | 0.05 degrees Celsius |
Max Drift Over Warranted Battery Life |
F20 and F40 series | 0.1 degree Celsius first two weeks after turn on
0.05 degree Celsius per six months thereafter |
D70 series | 0.05 degree Celsius first two weeks after turn on
0.05 degree Celsius per six months thereafter |
PhysioTel® Multiplus Implant Specifications |
Model | Weight
(gm) | Volume
(cc) | Minimum Animal Weight | Maximum Cage Size* | Battery Life Warranted (Months) |
F20-EET | 3.9 | 1.9 | 20
grams (SQ)
25 grams (IP) | 33 x 33 x 14
cm | 1.5 |
F40-EET | 7 | 3.7 | 175
grams | 42 x 42 x 18
cm | 3 |
F50-EEE | 11.5 | 5.5 | 175
grams | 42 x 42 x 18
cm | 2 |
D70-PCT | 49 | 33 | 2.5
kilograms | 1 x 1 x 1
meter | 3.5 |
D70-CCTP | 49 | 33 | 2.5
kilograms | 1 x 1 x 1
meter | 2.5 |
D70-PCTP | 49 | 33 | 2.5
kilograms | 1 x 1 x 1
meter | 2.5 |
D70-PCTR | 49 | 33 | 2.5 kilograms | 1 x 1 x 1 meter | 2.5 |
D70-EEE | 37 | 25 | 2.5
kilograms | 1 x 1 x 1
meter | 2 |
4ET-S1
and
4ET-S2 | 12.8 | 8.8 | 200
grams | 42 x 42 x 18
cm | 3
(user replaceable battery) |
Parameters Measured | Pressure, Biopotential, Temperature, Activity |
Outer Material | Silicone Elastomer |
Temperature Operating Range | 34 to 41 degrees Celsius |
Shelf Life | Battery - 1 year
Sterility - 1 year |
ON-Off Mechanism | Magnetically Actuated |
Receiver Compatibility | Any PhysioTel® RLA, RPC-1 or RMC-1 |
| |
Measurement Specifications |
Pressure Range | -20 to 300 mmHg |
Ambient Pressure Range | 670 to 800 mmHg |
Maximum Altitude | 625 meters |
| |
Pressure Accuracy |
Initial Accuracy | |
| |
Pressure Drift (D70 Series) | <2 mmHg per month nominal |
| |
Biopotential Channel Specifications | F20-EET,
F40-EET,
| D70-EEE,
F50-EEE,
4ET | Input Voltage Range | ±1.25mV | ±2.5mV | Channel Bandwidth | 1-50 Hz | 1-100 Hz | | | D70-CCTP, PCT, PCTP, PCTR | | | ±10mV | | | 1-100 Hz | |
Temperature Accuracy |
Initial Accuracy (D70-CCTP, D70-PCTP) | 0.35 degree Celsius |
Initial Accuracy (Others) | 0.25 degree Celsius |
Resolution | 0.05 degree Celsius |
Temperature Drift (D70-CCTP, D70-PCTP) | 0.2 degree Celsius per month |
Temperature Drift (F20-EET) | 0.15 degree Celsius per month |
Temperature Drift (Others) | First two weeks .1 degree Celsius
Thereafter .1 degree Celsius |
PhysioTel® F40-TT Implant Specifications |
Model | Weight
(gm) | Volume
(cc) | Minimum Animal Weight | Maximum Cage Size* | Battery Life Warranted (Months) |
F40-TT | 7.5 | 3.5 | 175
grams | 42 x 42 x 18
cm | 4 |
Temperature |
Temperature Operating Range | 22-41 degrees Celsius |
Initial Accuracy | 0.30 degree Celsius from 34-40 degree Celsius
0.45 degree Celsius all other |
Temperature Drift | 0.1 degree Celsius per 6 months |
Resolution | 0.01 degree Celsius |
| |
Battery Voltage | |
| Continuous monitoring via parameter |
Probe | |
| 1.8mm diameter
13cm thermistor-tipped probes |
| |
Specifications, performance and accuracy are valid for the warranted battery life and only when data are recorded using Dataquest A.R.T. with factory-supplied calibration values. Transmitters may work over wider ranges with reduced accuracy.
*An externally worn telemetry repeater is available to allow retransmission of the telemetry signals using multiple frequencies over a long range (3-10 m). This also permits the use of smaller transmitter models in large animals.
Request more information on a product.