| DSI PhysioTel® Transmitter Specifications |
PhysioTel® HD Transmitter Specifications
PhysioTel® TA Transmitter Specifications
PhysioTel® PA Transmitter Specifications
PhysioTel® EA, CA, ETA, and CTA Transmitter Specifications
PhysioTel® Multiplus Transmitter Specifications
| PhysioTel® HD Transmitter Specifications | |||||
Model |
Weight |
Volume |
Minimum Animal Weight |
Maximum Cage Size* |
Battery Life Warranted (Months) |
HD-S21 |
8 |
5.9 |
175 |
42 x 42 x 18 |
2 |
Specifications are nominal. Some parameters, such as minimum animal weight, may vary depending on research methodology and preference. For example, some researchers have implanted PhysioTel® HD transmitters in animals smaller than 175 grams. Larger cage sizes also can be accommodated with special configurations.
| Parameters Measured | 2 Pressures, Biopotential, Temperature, Activity |
| Outer Material | Silicone Elastomer |
| Temperature Operating Range | 34 to 41 degrees Celsius |
| Shelf Life | Battery - 1 year Sterility - 1 year |
| ON-Off Mechanism | Magnetically Actuated |
| Receiver Compatibility |
Any PhysioTel® RLA, RPC-1 or RMC-1 |
Measurement Specifications |
|
| Pressure Range | -20 to 300 mmHg |
| Catheter Length Options | 8, 10 or 15 cm |
| Ambient Pressure Range | 670 to 800 mmHG |
| Maximum Altitude |
625 meters |
Pressure Accuracy |
|
| Initial Accuracy | ± 3mmHg |
| Pressure Channel Drift |
<2 mmHg/month |
Biopotential Accuracy |
|
| Biopotential Input Range | ±5mV |
| Biopotential Input Bandwidth | 0.1-145 Hz |
| Biopotential Lead Length |
30cm, 60cm |
Temperature Accuracy |
|
| Initial Accuracy | 0.1 degree Celsius |
| Resolution | 0.05 degree Celsius |
| Temperature Drift |
First two weeks ≤ .1 degree Celsius Per 6 months after ≤ .1 degree Celsius |
Digital Data Information |
|
| Battery ON Time | Track study duration and device use; accurate within ±1.5 days at body temperature (i.e. ~37°C) |
| Animal ID (serial number) | Transmission of ID to assigned receiver prevents wrong data collection |
| Calibrations |
Simplifies configuration and ensures accurate calibration entry |
Specifications, performance and accuracy are valid for the warranted battery life and only when data are recorded using Dataquest™ A.R.T. 4.3, OpenART 4.3 or Ponemah 5.0 with factory-supplied calibration values. Transmitters may work over wider ranges with reduced accuracy.
*An externally worn telemetry repeater is available to allow retransmission of the telemetry signals using multiple frequencies over a long range (3-10m). This also permits the use of smaller transmitter models in large animals. See our website for more details.
| PhysioTel® TA Transmitter Specifications | |||||
Model |
Weight |
Volume |
Minimum Animal Weight |
Maximum Cage Size* |
Battery Life Warranted (Months) |
TA-F20 |
3.8 |
1.75 |
20 |
33 x 33 x 14 |
6 |
TA-F40 |
7.25 |
3.5 |
175 |
42 x 42 x 18 |
12 |
TA-D70 |
40 |
25 |
2.5 |
1 x 1 x 1 |
11 |
Specifications are nominal. Some parameters, such as minimum animal weight, may vary depending on research methodology and preference. For example, some researchers have implanted PhysioTel® TA transmitters in animals smaller than 20 grams. Larger cage sizes also can be accommodated with special configurations.
| Parameters Measured | Temperature and Activity |
| Outer Material | Silicone Elastomer |
| Temperature Operating Range | 34 to 41 degrees Celsius |
| Shelf Life | Battery - 1 year Sterility - 1 year |
| ON-Off Mechanism | Magnetically Actuated |
| Receiver Compatibility | Any PhysioTel® RLA, RPC-1 or RMC-1 |
Temperature Measurement Specifications |
|
| Initial Accuracy | 0.1 degree Celsius |
| Resolution | 0.01 degree Celsius |
| Max Drift Over | 0.05 degree Celsius first two weeks after turn on |
| Warranted Battery Life | 0.05 degree Celsius per six months thereafter |
Specifications, performance and accuracy are valid for the warranted battery life and only when data are recorded using Dataquest™ A.R.T. with factory-supplied calibration values. Transmitters may work over wider ranges with reduced accuracy.
*An externally worn telemetry repeater is available to allow retransmission of the telemetry signals using multiple frequencies over a long range (3-10m). This also permits the use of smaller transmitter models in large animals. See our website for more details.
| PhysioTel® PA Transmitter Specifications | |||||
Model |
Weight |
Volume |
Minimum Animal Weight |
Maximum Cage Size* |
Battery Life Warranted (Months) |
1.4 |
1.1 |
17 |
33 x 33 x 14 |
1.5 |
|
9.0 |
4.5 |
175 |
42 x 42 x 18 |
4 |
|
PA-D70 |
37 |
25 |
2.5 |
1 x 1 x 1 |
4 |
Specifications are nominal. Some parameters, such as minimum animal weight, may vary depending on research methodology and preference. For example, some researchers have implanted the PhysioTel® PA-C10 transmitter in mice smaller than 17 grams. Larger cage sizes also can be accommodated with special configurations. Minimum animal weight may vary based upon placement location, surgical skill and strain of animal used.
| Parameters Measured | Pressure and Activity |
| Outer Material | Silicone Elastomer |
| Temperature Operating Range | 34 to 41 degrees Celsius |
| Shelf Life | Battery - 1 year Sterility - 1 year |
| ON-Off Mechanism | Magnetically Actuated |
| Receiver Compatibility | Any PhysioTel® RLA, RPC-1 or RMC-1 |
Measurement Specifications |
|
| Pressure Range | -20 to 300 mmHg |
| Ambient Pressure Range | 670 to 800 mmHg |
| Maximum Altitude | 625 meters |
Pressure Accuracy |
|
| Initial Accuracy | ±3 mmHg |
Pressure Drift |
|
| PA-C40 | First month < 5 mmHg Second month < 4 mmHg Thereafter < 3 mmHg per month |
| PA-C10, PA-D70 | < 2 mmHg per month nominal |
Specifications, performance and accuracy are valid for the warranted battery life and only when data are recorded using Dataquest™ A.R.T. with factory-supplied calibration values. Transmitters may work over wider ranges with reduced accuracy.
*An externally worn telemetry repeater is available to allow retransmission of the telemetry signals using multiple frequencies over a long range (3-10m). This also permits the use of smaller transmitter models in large animals. See our website for more details.
| PhysioTel® EA, CA, ETA and CTA Transmitter Specifications | |||||
Model |
Weight |
Volume |
Minimum Animal Weight |
Maximum Cage Size* |
Battery Life Warranted (Months) |
EA-F20 |
3.9 |
1.9 |
20 |
33 x 33 x 14 |
4 |
ETA-F20 |
3.9 |
1.9 |
20 |
33 x 33 x 14 |
4 |
| ETA-F10 | 1.6 |
1.1 |
17 grams |
33 x 33 x 14 cm |
2 |
CA-F40 |
7.0 |
3.0 |
175 |
42 x 42 x 18 |
6 |
CTA-F40 |
7.0 |
3.0 |
175 |
42 x 42 x 18 |
6 |
CTA-D70 |
30.0 |
20.0 |
2.5 |
1 x 1 x 1 |
4 |
Specifications are nominal. Some parameters, such as animal weight may vary depending on research methodology and preference. For example, some researchers may have implanted PhysioTel CA and CTA transmitters in animals smaller than 175 grams.
| Parameters Measured | Biopotential, Temperature, Activity |
| Outer Material | Silicone Elastomer |
| Temperature Operating Range | 34 to 41 degrees Celsius |
| Shelf Life | Battery - 1 year Sterility - 1 year |
| ON-Off Mechanism | Magnetically Actuated |
| Receiver Compatibility | Any PhysioTel® RLA, RPC-1 or RMC-1 |
Temperature Measurement Specifications |
|
| Initial Accuracy | 0.1 degree Celsius |
| Resolution | 0.01 degree Celsius |
| ETA-F10 Initial Accuracy | 0.25 degrees Celsius |
| ETA-F10 Resolution | 0.05 degrees Celsius |
Max Drift Over Warranted Battery Life |
|
| F20 and F40 series | 0.1 degree Celsius first two weeks after turn on 0.05 degree Celsius per six months thereafter |
| D70 series | 0.05 degree Celsius first two weeks after turn on 0.05 degree Celsius per six months thereafter |
ETA-F10
|
Max Drift over Warranted Battery Life: 0.1 degrees Celsius per month |
Specifications, performance and accuracy are valid for the warranted battery life and only when data are recorded using Dataquest™ A.R.T. with factory-supplied calibration values. Transmitters may work over wider ranges with reduced accuracy.
*An externally worn telemetry repeater is available to allow retransmission of the telemetry signals using multiple frequencies over a long range (3-10m). This also permits the use of smaller transmitter models in large animals. See our website for more details.
| PhysioTel® Multiplus Transmitter Specifications | |||||
Model |
Weight |
Volume |
Minimum Animal Weight |
Maximum Cage Size* |
Battery Life Warranted (Months) |
3.9 |
1.9 |
20 |
33 x 33 x 14 |
1.5 |
|
7 |
3.5 |
175 |
42 x 42 x 18 |
3 |
|
7.5 |
3.5 |
175 |
42 x 42 x 18 |
4 |
|
F50-EEE |
11.5 |
5.5 |
175 |
42 x 42 x 18 |
2 |
C50-PXT |
11 |
6 |
175 |
42 x 42 x 18 |
2 |
D70-PCT |
49 |
33 |
2.5 |
1 x 1 x 1 |
3.5 |
D70-CCTP |
49 |
33 |
2.5 |
1 x 1 x 1 |
2.5 |
D70-PCTP |
49 |
33 |
2.5 |
1 x 1 x 1 |
2.5 |
49 |
33 |
2.5 kilograms |
1x1x1 meter |
2.5 |
|
D70-EEE |
37 |
25 |
2.5 |
1 x 1 x 1 |
2 |
12.8 |
8.8 |
200 |
42 x 42 x 18 |
3 |
|
Specifications are nominal. Some parameters, such as minimum animal weight, may vary depending on research methodology and preference. For example, some researchers have implanted PhysioTel® Multiplus transmitters in animals smaller than 175 grams. Larger cage sizes also can be accommodated with special configurations.
| Parameters Measured | Pressure, Biopotential, Temperature, Activity |
| Outer Material | Silicone Elastomer |
| Temperature Operating Range | 34 to 41 degrees Celsius |
| Shelf Life | Battery - 1 year Sterility - 1 year |
| ON-Off Mechanism | Magnetically Actuated |
| Receiver Compatibility | Any PhysioTel® RLA, RPC-1 or RMC-1 |
Measurement Specifications |
|
| Pressure Range | -20 to 300 mmHg |
| Ambient Pressure Range | 670 to 800 mmHg |
| Maximum Altitude | 625 meters |
Pressure Accuracy |
|
| Initial Accuracy | ±3 mmHg |
| Pressure Drift (C50-PXT) | First month < 5 mmHg Second month < 4 mmHg Thereafter < 3 mmHg per month |
| Pressure Drift (D70 Series) | <2 mmHg per month nominal |
Temperature Accuracy |
|
| Initial Accuracy (D70-CCTP, D70-PCTP) | 0.35 degree |
| Initial Accuracy (Others) | 0.25 degree |
| Resolution | 0.05 degree Celsius |
| Temperature Drift (D70-CCTP, D70-PCTP) | 0.2 degree Celsius per month |
| Temperature Drift (F20-EET) | 0.15 degree Celsius per month |
| Temperature Drift (Others) | First two weeks .1 degree Celsius Thereafter .1 degree Celsius |
Specifications, performance and accuracy are valid for the warranted battery life and only when data are recorded using Dataquest™ A.R.T. with factory-supplied calibration values. Transmitters may work over wider ranges with reduced accuracy.
*An externally worn telemetry repeater is available to allow retransmission of the telemetry signals using multiple frequencies over a long range (3-10m). This also permits the use of smaller transmitter models in large animals. See our website for more details.