DSI | Ponemah Digital Isolated Amplifier
Directly connect and effortlessly collect ECG, EMG, EEG and other biopotentials.
DSI’s versatile isolated amplifier is designed to
work with the DSI/Ponemah 7700 Series platform,
allowing researchers to collect signals from a broad array
of sources.
The 7700 platform provides optimal resolution for all
applications such as:
- Cardiovascular
- Respiratory
- Central Nervous System
- Hemodynamic
- Isolated Organ

Digital communication interface to DSI/Ponemah 7700 Platform.
Communication occurs through a digital interface to
communicate and update channel settings. Since the
7700 platform offers a digital interface, there is no
corruption in the data stream. The channel settings are
stored within the software and recalled via the software
protocol. This design feature also helps prevent undesired
changes to the signal conditioner since no knobs or
switches are necessary on the signal conditioner itself,
providing a high level of security and traceability.
Also, since the settings are saved electronically, signal
conditioner information can be recalled and reviewed
after the study has been concluded.

Isolated amplifier from DSI/Ponemah.
The DSI/Ponemah Isolated Amplifier is designed
specifically to work with signals from low level input
voltages with full scale range from 40µV to 40mV. All
channels can be independently configured. Each channel
has a programmable eight-pole low pass filter.
The total functionality of each channel is controlled through the
Ponemah data acquisition and analysis software platform.
Combine DSI hardware with industry-leading Ponemah
software for a robust, reliable and versatile platform trusted
by scientists in physiology, pharmacology and toxicology
laboratories worldwide.
At the core of DSI’s hardware platform are the
DSI/Ponemah 7700 digital signal conditioners and
accessories. A 7700 signal conditioner is a module that
plugs into the 7700 chassis which provides power to the
signal conditioner as well as provides a communication link to
the workstation.
DSI delivers better resolution of data using the latest digital
signal processing technology. 7700 signal conditioners filter
data, eliminating inaccuracies and signal offsets – preventing
issues that often arise from analog components. Each signal
conditioner is designed for a specific application(s) and ranges
in channel count from 3-32 channels per module making it a
more cost effective choice for users.
Additional Features of the Isolated Amplifier
- Industry standard 1.5mm jacks
- Long link cable to the 7700 chassis allows the pod to be placed next to the subject
- Ponemah data can be read and processed by DSI Neuroscore software for advanced
frequency analysis
- 12 differential pairs for collection of bio-potentials
- Ponemah software allows each channel to
have independent spans and ranges allowing
many types of signals to be collected at
one time.


Specifications
Input Circuit (typical of each channel)
- Differential isolated balanced to ground
- Greater than 10MΩ each input to common at DC
- Maximum input ±30VDC
- Leakage < 10uA at 265 VRMS
Measurement Range
Linear Range: 40µV to 40mV full scale
Signal Conditioner Output
- A/D Converter: 16 BITS
- Sample Rate: 20kHz per channel maximum
12 Bio-potential Pod Accuracies
Filters:
- Selectable: 100Hz, 300Hz or 1000Hz Low pass.
- Eight-pole DSP based Bessel settings of: 100Hz,
and 300Hz in conjunction with a hardware
implemented five-pole Bessel filter at 1000Hz.
- Non-selectable: .05Hz High Pass
(hardware implemented).
- Optional software filtering within Ponemah.
Common Mode: Less than -90dB
Crosstalk: Less than -60dB (measured at 36mVpk-pk
10 Hz with all channels set to same gain range)
Linearity: ±0.5% of full scale
Noise: Less than 20 uV pk-pk
Gain Error: Less than ±0.5% full scale
Zero Drift type: 0.01% of full scale per degree C
Request more information on this product.
|